Описание

Patch Clamp Recording Instruments with a choice of 3 Headstages are suitable for measuring currents in whole cell measurements and single channel studies. Microelectrodes or glass electrodes/micropipettes are not supplied. Other resistive-feedback headstages are also separately available for use with the Patch Clamp Amplifier.

Фото галерея

Технические характеристики

Independent V-hold and I-hold circuitry (3 operating modes)

V Clamp Mode: Input is clamped to a maximum of ±1 V. Commands are the sum of the holding voltage and any external input signals.

Zero Current Mode: Bridges the voltage and current clamp modes. All commands are inactive except the junction zero, which functions in this case as an offset control for the electrode and tip potentials associated with the pipette.

I Clamp Mode: In the current clamp mode, active commands are the current hold and any external command input signals, summed. Clamp stability is enhanced with the selection of three clamp speeds.

Compensation Controls

Voltage Offsets (including junction potentials): Automatically compensated with the auto zero or manually with the junction zero.

Capacity Compensation: Paired controls; fast (0-5 µs), medium (0-2 ms) and slow (0-20 ms) compensate the current transient caused by a step command signal. Each pair has a separate control for amplitude and time constant.

Series Resistance: Signal errors contributed by the electrode (access) resistance in whole cell recording are subtracted from the output with the series R comp control. The compensated resistance (0-100 MΩ) is read from a ten turn dial.

Leak Subtract: Used to compensate leakage current due to the voltage drop across the seal (shunt) resistance.

4-pole low-pass Bessel filter

Internally generated test signals

Variable duration zap circuit

Поделитесь ссылкой на эту страницу

Поделиться в vk
VK
Поделиться в facebook
Facebook
Поделиться в linkedin
LinkedIn
Поделиться в telegram
Telegram
favicon3

Похожее оборудование

A cardiac electrophysiology system designed to perform pacing and recording protocols that help to determin…
The Single and Dual Epithelial Voltage Clamp Systems are designed to be used with a corresponding Ussing Ch…
These systems are designed for two-electrode, whole-cell voltage clamping of large cells such as Xenopus oo…
Обратная связь
Мы свяжемся с вами в течение 15 минут
Наш менеджер подготовит для Вас коммерческое предложение и сразу свяжется с Вами.